Human Angiopoietin-like 4

Human Angiopoietin-like 4 DuoSet ELISA Summary

Assay   Type

Solid   Phase Sandwich ELISA

Format

96-well   strip plate

Assay   Length

4.5   hours

Sample   Type & Volume Required Per Well

100 μL

Sensitivity

0.068   ng/mL

Assay   Range

1.3 -   80 ng/mL (Cell Culture Supernates, Serum, Platelet-poor EDTA Plasma,   Platelet-poor Heparin Plasma, Urine, Human Milk)

Specificity

Natural   and recombinant human Angiopoietin-like4

 

Cross-reactivity

<   0.5% cross-reactivity observed with available related molecules.< 50%   cross-species reactivity observed with species tested

Interference

No   significant interference observed with available related molecules.

Product Summary


Shipping

The   product is shipped at ambient temperature. Upon receipt, store it immediately   at the temperature recommended below.

Storage

Store   the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: Angiopoietin-like Protein 4/ANGPTL4

Angiopoietin-like, also known as angiopoietin-related proteins, are structurally related to the angiopoietins. They are secreted proteins with N-terminal coiled-coil domains and C-terminal fibrinogen-like domains. Unlike the angiopoietins, ANGPTLs do not bind Tie-2. Seven ANGPTL proteins have been identified (ANGPTL1 through 7).

Long   Name:

Angiopoietin-like4

Entrez   Gene IDs:

51129   (Human); 57875 (Mouse); 362850 (Rat)

Alternate   Names:

Angiopoietin   like Protein 4; angiopoietin-like 4; Angiopoietin-like Protein 4; ANGPTL4;   ARP4fasting-induced adipose factor; FIAF; FIAFhepatic angiopoietin-related   protein; Hepatic fibrinogen/angiopoietin-related protein; HFARP;   HFARPANGPTL2; NL2; peroxisome proliferator-activated receptor (PPAR) gamma   inducedangiopoietin-related protein; PGAR; PGARangiopoietin-related protein   4; pp1158; PPARG angiopoietin related protein

Assay Procedure

Refer to the product datasheet for the complete assay procedure.

Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate.

1.     Prepare all reagents, standard dilutions, and samples as directed in the product insert.

2.     Remove excess microplate strips from the plate frame, return them to the foil pouch containing the desiccant pack, and reseal.

3.     Add 50 μL of Assay Diluent to each well.

4.     Add 50 μL of Standard, control, or sample to each well. Cover with a plate sealer, and incubate at room temperature for 2 hours.

5.     Aspirate each well and wash, repeating the process 4 times for a total of 5 washes.

6.     Add 100 μL of Conjugate to each well. Cover with a new plate sealer, and incubate at room temperature for 2 hours.

7.     Aspirate and wash 5 times.

8.     Add 100 μL Substrate Solution to each well.

9.     Add 100 μL of Stop Solution to each well. Read at 450 nm within 30 minutes. Set wavelength correction to 540 nm or 570 nm.